41WIKA operating instructions, model CPH7600
14051628.01 11/2012 GB/D
GB
7. Maintenance, cleaning, and recalibration
7. Maintenance, cleaning, and recalibration
7.1 Maintenance
This hand held pressure calibrator is maintenance-free.
Repairs must only be carried out by the manufacturer.
CAUTION!
To avoid personal injury or damage to the calibrator, only use
accessories supplied by WIKA and ensure that no water nds its way into
the case.
7.2 Cleaning
CAUTION!
Before cleaning, vent and switch o the hand-held pressure
calibrator.
In order not to damage the display or case, do not use any solvent or
scouring agents for cleaning.
Clean the instrument with a moist cloth.
Electrical connections and pressure connections must not come into
contact with moisture.
For information on returning the instrument see chapter 9.2 "Return".
7.3 Recalibration
DKD/DAkkS certicate - Certicates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with
time intervals of approx. 12 months. Every factory recalibration includes, additionally,
an extensive free-of-charge check of all system parameters with respect to their
compliance with the specication. The basic settings will be corrected if necessary.