WIKA operating instructions dierential pressure transmitter, model DPGT40 17
14093267.01 03/2014 GB/D/F/E
6. Commissioning, operation / 7. Options and accessories
6.3 Setting up a voltage supply
The voltage supply is made via a power supply unit or a control unit which
provides the energy limitation.
The power supply for the pressure transmitter must be made via an energy-
limited electrical circuit in accordance with section 9.3 of UL/EN/IEC 61010-1, or
an LPS to UL/EN/IEC 60950-1, or class 2 in accordance with UL1310/UL1585
(NEC or CEC). The power supply must be suitable for operation above 2,000 m
should the pressure transmitter be used at this altitude.
7. Options and accessories
7.1 4-way valve manifold
Isolation of the ⊕ and ⊖ process lines for removing or testing the measuring
instrument without interrupting the running process operation.
Protection of the instrument against
excessive overpressure loading, such as
in pressure tests and undened operating
conditions (including intermittent shutdown).
Pressure compensation for zero point
checking with running processes, and
avoiding one-sided overpressure loading
during start-up and operation phases (with
opened pressure compensating valve).
Venting the measuring lines with liquid
media and ushing of the measuring lines,
in order to remove contamination.
Vent valve
Shut-o valve
⊖ side
Shut-o valve
⊕ side
Pressure
compensating valve
Specications for handling
Sequence of operations to start measurement
1. Open the pressure compensating valve (middle valve spindle)
2. Open the shut-o valve for the negative media chamber (⊖, right-hand valve)
and the positive media chamber (⊕, left-hand valve)
3. Close the pressure compensating valve
Sequence of operations to ush/vent the measuring lines
1. Start: Open the shut-o valve for the ⊕ and ⊖ media chamber, open the
pressure compensating valve and vent valve
2. Finish: Close the pressure compensating valve and vent valve