WIKA operating instructions level sensor model FFG
13445627.01 12/2010 GB/D/F
8. Technical specications
Type of protection, explosion and temperature class
II 1/2 G EEx ia IIC T6–T2 II 2 D T95 °C IP 6X
II 1/2 G EEx ia IIB T6–T2 II 2 D T95 °C IP 6X
(pharmaceutical design)
Equipment group: II
Equipment category: 1G (oat and guide tube)
2G / 2D (connecting case)
Type of protection: EEx ia
Explosive sub-group: IIC or IIB (drop protection PTFE)
Temperature class: T2, T3, T4, T5, T6
Summary of electrical specications of authorized design
variants
Power supply:
Voltage Consumption
of power
Inductivity
(extern.
eective)
Capacity
(extern.
eective)
Type of
protection
DC 10 ... 30 V max. 700 mW max. 250 μH max. 5 nF EEx ia
Permissible data of control unit:
Ui Ii Pi Li Ci Type of protection
≤ 30 V ≤ 200 mA ≤ 1 ≥ 250 μH ≥ 5 nF EEx ia
8. Technical specications