Thermowell
Stem
Required safety
clearance
14004105.01 05/2011 GB/D/F/E
WIKA operating instructions model 54, Twin-Temp 11
6. Commissioning, ... / 7. Maintenance and ...
CAUTION!
When using thermowells, please
ensure that the stem does not
touch the bottom of the thermowell
since, due to the dierent
expansion coecients of the
materials, the stem may become
buckled at the bottom of the
thermowell.
(Formula for the calculation of the
stem length l
1
see the respective
thermowell‘s data sheet)
7. Maintenance and cleaning
7.1 Maintenance
The combined bimetal thermometer is maintenance-free!
The indicator should be checked once or twice every year. To do this
the instrument must be disconnected from the process and checked
using a temperature calibrator.
Repairs must only be carried out by the manufacturer.
7.2 Cleaning
CAUTION!
Clean the thermometer with a moist cloth.
Wash or clean the disconnected thermometer before
returning it in order to protect personnel and the
environment from exposure to residual media.