4.4 General Specifications
Size 14.02” wide x 7.55” high x 12.42” deep (35.61 cm x
19.2 cm x 31.55 cm). See Figure 5.2.
Weight 36 lbs. (16.33 kg) with all internal options.
Power Input Requirements 100-240 VAC, 47-63 Hz, 75 VA max.
Fuses: 1.5A, 230V, Type T LITTLEFUSE 31301.5
Digital I/O Pin 1: +5 VDC, 100 mA max.
Pin 2: Input TTL levels. Voltage input should be limited
to > -0.3 VDC and < +5.3 VDC.
Pin 3: Output pulled up to 5 VDC with a 10k ohm
resistor. Source limited to 450uA, Sink 10mA
Pin 4: Ground.
Pneumatic Interfaces 7/16” – 20 female SAE.
1/8” FNPT adapters provided.
Particle Filters The instrument has 20-micron filters on all pressure
ports through the manifold.
The barometric transducer has no filters.
Pneumatic Overpressure Protection Sensors are protected by relief valves.
Compensated Temperature Range 15°C to 45°C.
Operating Temperature Range 0°C to 50°C. Note: This is not the compensated
temperature range.
Storage Temperature Range 0°C to 70°C. Minimal vibration.
Non-condensing humidity.
Local User Interfaces 8.4” color LCD display with 8 wire resistive touch
screen.
Remote User Interfaces RS-232, Ethernet, and IEEE-488.1 are standard.
Warm-up Approximately 15 minutes to achieve full accuracy
depending on environment.
Reading Rate Typically 32 readings per second.
Measure Response Time <333 mS for FS step.
Orientation Effects Negligible, can be removed with re-zeroing.
Operating Environment 5 to 95% RH non-condensing.
Pressure Media Clean, dry, non-corrosive, non-combustible, non-
oxidizing gases. Not suitable for oxygen use.
Compliance
See section “1.3 Compliance”
Options Transport Case, Rack Mount Kit, Calibration Sled
Kit, Secondary Transducer, Barometric Reference
Transducer, Virtual Delta Channel