9037926 03/2005
Temperature Dry Well Calibrators CTD 9300
3. Moduses and Functions
GB
40 WIKA Operating Instructions CTD 9300 Rev. 1
This menu is used to set the control parameters.
3.1.3.3 Controler
Fig. 25: Control parameter set
The parameter set in the basic configuration is as follows (device-
dependent):
Control parameter CTD 9300-650 CTD 9300-165
P0 1.00000E+00 1.00000E+00
P1 1.00000E-03 1.00000E-02
P2 1.00000E+00 1.00000E+00
P3 0.00000E+00 0.00000E+00
P4 1.00000E-01 1.00000E-01
Where:
Control parameter Designation
P0 internal amplification
P1 adjustment of the disturbance variable
P2 time constant
P3 test parameter
P4 safety limit of the dynamic
Warning!
Ex works, the control parameters are set for an optimum
control behaviour.
The user should not edit these settings.