9037926 03/2005
Temperature Dry Well Calibrators CTD 9300
3. Moduses and Functions
GB
48 WIKA Operating Instructions CTD 9300 Rev. 1
Designation
Probe
t-open
t-close
Hysteresis
Information
Type of probe
opening temperature
of the switch
closing temperature
of the switch
The difference between
"t-open" and "t-close"
Note
Temperature switch
in this example: 107.32 °C
in this example: 103.98 °C
in this case: 3.34 K
The following information is provided:
3.3.2.2 Calibration Menu “TC-Probe”
In this menu, the output signal
of a thermocouple can be
displayed in addition to the
nominal and actual
temperatures of the calibrator.
Fig. 35: Calibration menu
“TC-probe”
The following information is provided:
Designation
Probe
t-Probe
Error
Signal
Information
Type of probe
Temperature of the
probe
Error of the probe
in K
Electric output signal
of the probe
Note
Thermocouple: TCJ-3, i.e.
Thermocouple J at input 3
in this example: 102.67 °C
Deviation from the int. (ext.)
reference thermometer
In the case of the thermocouple,
the thermal voltage in µV