Wika T12 Bedienungsanleitung Seite 28

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 72
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 27
GB
3211517.0707/2010GB/D
28 WIKAoperatinginstructionstemperaturetransmitterT12
8.1 Model overview and their European approvals
Model Ex protection and
Approval No.
Ignition
protection type
head
mounting
version
rail
mounting
version
T12.1x.xx2 T12.30.xx2
II1GEExiaIIB/IICT4/T5/T6
DMT98ATEXE008X
intrinsicallysafe
equipment
T12.10.xx9 T12.30.xx9
II3GExnLIICT4/T5/T6
energy-limited
equipment
II3GExnAIICT4/T5/T6
non-sparking
equipment
II3GExicIICT4/T5/T6
intrinsicallysafe
equipment
8.2 Special conditions for safe use
T12.30.xxx:
Thesurfaceofthecaseisnotconductive.Thetemperaturetransmitter
mustbemountedinawaythatelectrostaticchargeswillnotoccur.
T12.xx.xx2:
Temperaturetransmittersinahazardousareaaresuppliedonly
withassociatedintrinsicallysafeapparatusthatareapprovedfor
thishazardousarea.Thesetransmittersmustbemountedinacase
thatmustatleastcorrespondtofollowingingressprotectionIP20
accordingtoEN60529/IEC60529.
T12.10.xx2 for category II 1G/IIC in addition:
Thesurfaceofthecaseisnotconductive.Thetemperaturetransmitter
mustbemountedinawaythatelectrostaticchargeswillnotoccur.
T12.xx.xx9 (Use as energy-limited equipment II 3G EEx nL):
Thesupplycurrentcircuitmustfulltherequirementsforignition
protectiontypeII3GEExnLenergy-limitedperEN50021.These
temperaturetransmittersmustbemountedinacasethatmustat
leastcorrespondtofollowingingressprotectionIP54accordingto
EN60529/IEC529.
8. Notes for mounting and operating in ...
Seitenansicht 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 71 72

Kommentare zu diesen Handbüchern

Keine Kommentare