Wika Pascal ET Bedienungsanleitung Seite 31

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 88
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 30
6 Commissioning, operation
WIKA Operating Instruction, Model Pascal ET and Pascal ET/IS 31
GB
Once a particular parameter is selected, for example pressing T” for temperature, the instrument
asks for the channel role:
Figure 13 Channel role selection
NONE to use the channel for simple measurement
REF to use the channel as reference in a comparison calibration
DUT to use the channel as measurement of a Device Under Test in a comparison
calibration
Press one of the above button (NONE, REF, DUT) to define the channel role.
Pressing REF or DUT the procedure continues as shown below:
I/O to assign a particular I/O module to the software channel.
KEYBOARD to setup by keyboard a particular value read from an external instrument
(i.e. analogue gauge).
Figure 14 Channel mode selection
Seitenansicht 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 87 88