Wika Pascal ET Bedienungsanleitung Seite 52

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 88
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 51
6 Commissioning, operation
52 WIKA Operating Instruction, Pascal ET and Pascal ET/IS
GB
Calculation is possible only when channel 1 is assigned for mV measurement (bridge signal) and
channel 2 is assigned for V generation, range 0÷20 V (bridge supply). Moreover channel 1 should
not be set in error mode.
In this case, below what will be displayed:
Figure 60 Cell Load calculation
Pascal ET or Pascal ET/IS allows to manage the sensitivity of the load cell from 0 mV/V up to
9.99999 mV/V.
Moreover, on channel 4, the SCALING function can be assigned through the menu CHANNEL in
order to display an engineering unit more appropriate (for example Kg or a pressure unit).
Figure 61 Cell Load calculation with Scaling function active
Seitenansicht 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 87 88

Kommentare zu diesen Handbüchern

Keine Kommentare