Application (see page 12):
Example 1
Sensor type Pt100, congured measuring range = -50 … +100 °C, so congured
measuring span = 150 K.
This is not smaller than 84 K. Thus the total safety accuracy is 2 % FS,
so 2 % * 150 K = 3 K, or 2 % * 16 mA = 320 μA in terms of the current output
Example 2
Sensor type Pt100, congured measuring range = 0 … 50 °C, so congured
measuring span = 50 K
This is smaller than 84 K, thus the total safety accuracy is 2 K,
thus 2 K / 50 K = 4 %, and 4 % * 16 mA = 640 μA in terms of the current output
2.7 Conguration changes
WARNING!
During the conguration change, the safety function is not active! Safe
operation is only admissible with activated write protection (password).
Carry out conguration changes within the permissible specications in accordance
with chapter 2.1 “Intended use in safety applications”.
With the supplied conguration tools, such items as the write protection for model
T32.xS can be set:
Conguration software WIKA_T32
AMS
SIMATIC PDM
DTM in conjunction with operating software to the FDT/DTM Standard, e.g.
PACTware, FieldMate
HART
®
hand-held terminal FC475, FC375, MFC4150
WARNING!
The safety function must be checked by testing following any conguration
procedure.