55WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
a
b
c
d
e
f
g
h
i
..ReferenceSensor
Reference sensor
Sensor no:
R.-Start:
Range-End:
Unit:
Class:
Pressuretype:
Overload:
362A
0.00
10.00
bar
0.025
rel
0 sec
%FS
Read
Ref-Value:
Offset:
Cal-Dat.:
0.001
0.000
2011/05/10
7.4.3 SETUP additional menu items: ReferenceSensor
7. Commissioning, operation
a) Sensor no:
Sensor number of the currently connected CPT6000 reference pressure sensor
b) R.-Start/Range-End:
Start of measuring range and end of measuring range of the currently connected
CPT6000 reference pressure sensor
d) Unit:
Basic pressure unit of the CPT6000 reference pressure sensor
d) Class:
Accuracy of the measuring chain of the CPH6000 with connected CPT6000 refer-
ence pressure sensor
e) Pressuretype:
Pressure type of the currently connected CPT6000 reference pressure sensor
(Gauge/relative pressure or absolute pressure)
f) Overload:
Length of time for which the CPT6000 reference pressure sensor was unaccept-
ably overloaded.
If the value here is not equal to zero, then it is highly probable that
the instrument no longer meets its specied accuracy class. The only
solution for this is an immediate recalibration. (For absolute pressure
sensors < 1 bar this function is deactivated, since for this measuring
range atmospheric pressure already represents an overload)