6 Commissioning, Operation
16 WIKA Operating Instruction, Model CPH6300
Set Output: Universal output adjustments
Interface and analog output off
Serial interface activated
Analog output activated
Base address for serial interface communication
(only at Out = SEr)
DIF
CH
Channel that should be used for analog output
(only at Out = dAC)
e.g.-5.00..5.00 mbar
Zero-point setting for Out = dAC: Measuring value which
should correspond to output 0 V
e.g.-5.00..5.00 mbar
Slope setting for Out = dAC: Measuring value which should
correspond to output 1 V
Set Corr: Input adjustment (1-channel version)
Zero adjustment/offset of sensors
oFF
No zero adjustment of sensors
sensor dep.
e.g.-5.00..5.00 mbar
The offset of sensor will be displaced by this value to
compensate for deviations of the probe or of the measuring
Slope adjustment of sensors
No slope adjustment of sensor
-2.000 ... 2.000
The slope of sensor will be changed by this factor [%] to
compensate deviations of probe or measuring device
Set Corr: Input adjustment (2-channel version)
Zero adjustment/offset of sensors 1
oFF
No zero adjustment of sensors 1
e.g.-5.00..5.00 mbar
The offset of sensor 1 will be displaced by this value to
compensate for deviations of the probe or of the measuring
Slope adjustment of sensors 1
oFF
No slope adjustment of sensor 1
-2.000 ... 2.000
The slope of sensor 1 will be changed by this factor [%] to
compensate deviations of probe or measuring device
Zero adjustment/offset of sensors 2
oFF
No zero adjustment of sensors 2
e.g.-5.00..5.00 mbar
The offset of sensor 2 will be displaced by this value to
compensate for deviations of the probe or of the measuring
Slope adjustment of sensors 2
No slope adjustment of sensor 2
-2.000 … 2.000
The slope of sensor 2 will be changed by this factor [%] to
compensate deviations of probe or measuring device