Wika CTM9100-150 Bedienungsanleitung Seite 26

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 100
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 25
26 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
7. Operating elements of the calibrator/micro calibration bath
7.4.2 Switching on automatic control
For certain tasks, it is advantageous to switch o control (e.g.
to make settings on the calibrator or micro calibration bath).
In calibration mode, press button P for approx. 5 secs., this
opens the main menu.
The upper PV display will show OPEr.
In the lower SV display, the LED SET will blink.
Press button 5 or button 6 until OFF is displayed.
Conrm with button P.
On the upper PV display, the current reference temperature
and OFF are indicated alternately.
On the lower SV display, the currently-selected set tempera-
ture is displayed.
CAUTION!
The control is now switched o and the
reference temperature will fall steadily,
unless it is corrected.
7.4.3 Switching on automatic control
The control is switched o when the following display is
shown:
On the upper PV display, the current reference temperature
and OFF are indicated alternately.
On the lower SV display, the currently-selected set tempera-
ture is displayed.
The control is switched back on by pressing button P for
approx. 5 secs., this opens the main menu.
The upper PV display will show rEG.
In the lower SV display, the LED SET will blink.
Conrm switching the control on by pressing button P.
CAUTION!
Control is now activated. The calibrator or
micro calibration bath will default to calibra-
tion mode and the set temperature will be
reached.
Display in main menu
Control OFF menu
Display with setting Control OFF
Display rEG
Display with setting Control OFF
Seitenansicht 25