GB
4. Design and function
14083458.01 09/2013 GB/D
WIKA operating instructions models TR12, TC12 (Ex i) 13
Vibration resistance at the sensor tip
Resistance thermometer
6 g peak-to-peak, wire-wound measuring resistor or metal thin lm (standard)
20 g peak-to-peak, wire-wound measuring resistor or metal thin lm (option)
50 g peak-to-peak, metal thin lm measuring resistor (option)
1)
1) For measuring insert diameter < 8 mm
Thermocouples
50 g, peak-to-peak
Electrical connection
On the connection side, the thermometer has a housing or bare wire ends. The housing
design will contain the terminals or certied transmitters. Optionally, separately certied
digital displays can be built into the housing.
The connection side of the measuring insert consists of a transition sleeve with connected
connection wires.
CAUTION!
Thermometers of this model range must when used in zone 0 be installed with a
thermowell (minimal wall thickness: 1 mm) The design of the thermowell can be
selected as desired, but the operational process data (temperature, pressure,
density and ow rate) must be taken into account.
Possible sensor measuring ranges:
Model TR12: -200 ... +600 °C
Model TC12: -40 ... +1,200 °C
The following mounting and operating information has been compiled with care.
However, it is not possible to consider all potential applications.