
l primary_value
- Process Value after min/max-adjustment and linearization. Selected as input to AIFB by setting
'Channel' = 1. Unit derives from 'Primary_value_unit'
l primary_value_unit
- Unit code of 'Primary_value'
- %
l secondary_value_1
- Process pressure. Selected as input to AIFB by setting 'Channel' = 2. Unit derives from
'Secondary_value_1_unit'
l secondary_value_1_unit
- Unit code of 'Secondary_value_1'
- bar, PSI, …, m, ft, …; in case of length type engineering unit and access to parameters the
corresponding values will be converted by density factor
l secondary_value_2
- Value after min/max-adjustment. Selected as input to AIFB by setting 'Channel' = 3. Unit derives
from 'Secondary_value_2_unit'
l secondary_value_2_unit
- Selected unit code for "secondary_value_2"
l sensor_value
- Raw sensor value, i.e. the uncalibrated measurement value from the sensor. Unit derives from
'Sensor_range.unit'
l sensor_range
- "Sensor_range.unit' refers to 'Sensor_value', 'Max/Min_peak_sensor_value', 'Cal_point_hi/lo'
- includes sensor unit: bar, PSI …; only unit part of DS-68 is writable
l simulate_primary_value
l simulate_secondary_value_1
l simulate_secondary_value_2
l device status
- "0: ""OK"" 13: ""non-specific error"" 17: ""Cal span too small"" 34: ""EEPROM memory fault"" 36:
""ROM memory fault"" 37: ""RAM memory fault"" 40: ""non-specific hardware fault"" 41: ""Sensor
element not found"" 42: ""No leaking pulse"" 43 : ""No trigger signal"" 44: ""EMI error"" 113:
""Communication hardware fault"""
l linearization type
- Possible types of linearization are: linear, user defined, cylindrical lying container, spherical
container
- "0: ""Linear"" 1: ""User def"" 20: ""Cylindrical lying container"" 21: ""Spherical container"""
l curve_points_1_10
- X and Y values for the user defined linearization curve
l curve_points_11_20
- X and Y values for the user defined linearization curve
l curve_points_21_30
- X and Y values for the user defined linearization curve
l curve_points_31_33
- X and Y values for the user defined linearization curve
l curve status
Differential pressure transmitter DPT10 - metal measuring diaphragm • Foundation Fieldbus 83
11 Supplement
37246-EN-100120
Kommentare zu diesen Handbüchern