2.3 Additional safety instructions for instruments per ATEX
WARNING!
Non-observance of these instructions and their contents
may result in the loss of explosion protection.
WARNING!
Observe the applicable regulations for the use
of Ex-class instruments (e. g.: EN 60079-0:
2006, EN 60079-11: 2007, EN 60079-26: 2007,
EN 60079-27: 2006 + 2008, EN 61241-0: 2006,
EN 61241-11: 2006, EN 60079-15: 2005).
Do not use transmitters with any damage to the exterior!
Do not congure the transmitter, programming unit
(PU-448) and PC in hazardous areas!
2.4 Special hazards
WARNING!
Observe the information given in the applicable type
examination certicate and the relevant country-
specic regulations for installation and use in potentially
explosive atmospheres (e.g. IEC 60079-14, NEC, CEC).
Non-observance can result in serious injury and/or
damage to the equipment.
For additional important safety instructions for instruments
with ATEX approval, see chapter 8 “Information on
mounting and operation in hazardous areas (Europe)”.
WARNING!
For hazardous media such as oxygen, acetylene,
ammable or toxic gases or liquids, and refrigeration
plants, compressors, etc., in addition to all standard
regulations, the appropriate existing codes or regulations
must also be followed.