23WIKA operating instructions, model CEP3000
14036068.02 08/2012 GB/D
GB
8. Faults / 9. Return and disposal
8. Faults
Display Type of fault Measures
Low battery power, functioning is only
guaranteed for a short period of time
Insert new batteries
OL
-OL
Reading is signicantly above or below
the measuring range
Test: Is the measurement parameter
within the approved measuring range
of the instrument?
No display
or unden-
able characters,
instrument is not
responding to
key press
Batteries are discharged Insert new batteries
System error Switch o the instrument, wait for a
short period of time, switch on again
Instrument faulty Send in for repair
CAUTION!
If faults cannot be eliminated by means of the measures listed above,
the hand-held temperature calibrator must be shut down immediately,
and it must be ensured that signal is no longer present, and it must be
prevented from being inadvertently put back into service.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
9.1 "Return".
9. Return and disposal
WARNING!
Residual media on the dismounted hand-held temperature calibrator
can result in a risk to persons, the environment and equipment. Take
sucient precautionary measures.