3.2 Transmitter with model PGT23.063
Power supply U
B
DC V 12 U
B
30≤
Supply voltage effect % FS/10 V 0.1
Permissible residual ripple % ss 10
Output signal, variant I 4 … 20 mA, 2-wire
Permissible max. load R
A
R
A
(U
B
- 12 V)/0.02 A with R
A
in and U
B
in Volt, but
max. 600
Effect of load % FS 0.1
Linearity % of span 1.0 (limit point calibration)
1)
Output signal accuracy % FS < 1.6 electronic
1)
Input signal rotational angle < ° 0 … 270
Expanded special span < ° < 330 (option)
Long-term stability of electronics % FS/a < 0.5
Permissible ambient temperature °C -40 … +60
Permissible storage temperature °C -40 … +70
Permissible medium temperature °C -40 … +100
CE-conformity 2007/108/EC interference emission (Limit Class B) and
immunity to EN 61 326-1
Wiring Free cable or as option miniature plug connector M8 x 1,
4-pin
Wiring protection IP 54 per EN 60 529 / IEC 529, filled IP 65
Connection details see page 9
1) For technical reasons, up to the first scale marking, the measured value can lie outside of the class accuracy.
3.3 Model PGT23.063
Working pressure
- Steady 3/4 x full scale range
- Fluctuating 2/3 x full scale range
- Transient Full scale range
Temperature influence
max. ± 0.4 % of full scale range/10 K (when the temperature of the
measuring system deviates from +20 °C reference temperature)
IP protection class
- Housing IP 54 per EN 60 529 / IEC 60 529
- With liquid filling IP 65
Permissible temperatures
- Ambient filled/unfilled -40 ... +60 °C
- Medium Maximum +100 °C
Further technical data see data sheet PV 12.03
)
)
3. Technical data