21WIKA operating instructions, model CPH62I0
11221801.02 02/2011 GB/D/F/E
GB
6. Commissioning, operation
Menu Param. Setting Meaning
´MENU´
▶ ▲ or ▼
SEt
ConF
Set Conguration: Generic settings
Unit
mbar, bar, ... Unit: Unit of display *, **
SL
oFF / on Sea-level correction: on or o *, **
Alti
-200 ... +9999
Altitude: Input of altitude above sea-level [m] (only if on) *, **
rAtE
Rate: Measuring rate (see chapter 6.3.3) *
Slo Slow: Measuring rate (4 Hz ltered, low power consumption) *
FASt Fast: Measuring rate, ltered (> 1000 Hz) *
P.dEt Peak detection: fast measuring rate, unldered (> 1000 Hz) *
t.AVG
1 ... 120 Averaging interval in seconds, used by the averaging function o
oFF Averaging function deactivated
P.oFF
1 ... 120 Auto Power-O delay in minutes. Device will automatically switch
itself o if, during this interval, no key is pressed, or there is no
serial communication.
oFF Auto Power-O function inactive (continuous operation)
Out
oFF Function of the output: No output function, lowest power consump-
tion
SEr Output is serial interface
dAC Output is analogue output 0 ... 1 V
Adr.
01, 11 ... 91 Communications address of the interface
dAC.
CH1, CH2 or
CH DIF
Choice of the input to be the source for the analogue output (if Out
= dAC)
dAC.0
eg. -5.00 ...
+5.00 mbar
Enter desired value at which the analogue output potential should
be 0 V (if Out = dAC)
dAC.1
eg. -5.00 ...
+5.00 mbar
Enter desired value at which the analogue output potential should
be 1 V (if Out = dAC)
SEt
CAL
Set Calibration: Adjustment of sensors
OFS.1
Sensordep.,
e.g. -5.00 ...
+5.00 mbar
The offset of sensor 1 will be displaced by this value to compen-
sate for deviations in the probe or in the measuring device.
OFF Zero displacement inactive (=0.0 °)
SCL.1
eg. -5.00 ...
+5.00 mbar
The measuring scale of sensor 1 will be changed by this factor
[%] to compensate deviations of probe or measuring device.
OFF Scale correction factor inactive (=0.000)
OFS.2
Sensordep.,
e.g. -5.00 ...
+5.00 mbar
The offset of sensor 2 will be displaced by this value to compen-
sate for deviations in the probe or in the measuring device.
OFF Zero dispacement deactive (=0.0°)
SCL.2
eg. -5.00 ...
+5.00 mbar
The measuring scale of sensor 2 will be changed by this factor
[%] to compensate deviations of probe or measuring device.
OFF Scale correction factor inactive (=0.000)