33WIKA operating instructions, model CPH62I0
11221801.02 02/2011 GB/D/F/E
GB
8. Faults
Display Meaning Measures
Err.1
Measured value above allowable
range
Check: Is the pressure outside the
sensor‘s range? -> measured value
too high!
Sensor faulty Return to manufacturer for repair
Err.2
Measured value below allowable
range
Check: Is the pressure outside the
sensor‘s range? -> measured value
too low!
Sensor defective Return to manufacturer for repair
Err.3
Display range exceeded Check: value above 9999 -> too high
to be displayed!
Err.4
Display range underow Check: value below ‚-2000‘ (Tare?)
-> too low!
Err.7
System error Return to manufacturer for repair
Err.11
Value could not be calculated Choose dierent unit
Calculation range exceeded Choose dierent unit
WARNING!
If faults cannot be eliminated by means of the measures listed above,
the instrument must be shut down immediately, and it must be ensured
that pressure and/or signal are no longer present, and it must be
prevented from being inadvertently put back into service.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter “9.2
Return”.